Produit

2
3

TFD04 Split Style Syringe Pump

brève description:

Plage de débit:

Nombre maximum de canaux: 8

Envoyez-nous un e-mail

Introduction

Lead Fluid TFD04 laboratory syringe pump, transmission accuracy error<±0.5%, can independently control 1~8 channel operation. Split design, more flexible usage. 7-inch HD color touch screen display and operation, fonctions puissantes, opération simple, rich operating information. Communication RS485, prend en charge le protocole MODBUS.

Fonctionnalité

• Can independently control 1~8 channel operation.
• 7-inch HD color touch screen display and operation, bringing a good user experience.
• Système logiciel LF-Touch-OS, friendly interface, reliable operation.
• Accurate flow control, transmission accuracy error<±0.5%.
• Intelligent flow correction function.
• Quatre modes de travail:only infuse, only withdraw, infuse/withdraw, withdraw/infuse.
• With fast filling and emptying function.
• Support start delay and cyclic operation.
• Save five groups of formula parameters.
•Communication RS485, Protocole MODBUS, support the setting of communication parameters, convenient to connect with various control device.
• External control signal controls start, stop, pause, and physical isolation.
• Super anti-interference characteristics, conception à large tension, adapté à un environnement d'alimentation complexe
• High-strength metal case with protective coating, corrosion-resistant and easy to clean.
• The internal circuit is treated with special protection to prevent dust, moisture and static electricity.

Paramètre technique

Canal 1~8 channels,independently control
Syringe size 0.5μL~60mL(replace the execution unit)
Syringe selection built-in mainstream brand syringe specifications, support customization
Working mode infuse, withdraw ,infuse/withdraw, withdraw/infuse
Précision error<±0.5%(>30% of pump stroke)
Flow calibration intelligent guide
Afficher 7 inch 65565 color LCD
Operation method Touch screen
Delay time 0.1~999 s/min/h
Cycles 0~9999
Cycles interval time 0.1~999 s/min/h
Power-down memory Memorize parameters set before power off
Status signal output 5V TTL,1 road indicates running status, 1 road indicates direction status
Control signal input 5~24V Wide voltage, control start, stop and pause
Communication RS485,support Modbus protocol
Source de courant CA 100 ~ 240 V,50Hz/60 Hz
Consommation d'énergie Controller 40W+30W (each execution unit)
Environnement temperature5~40℃ ,humidité relative <80%
Controller dimension 200×142×88mm
Controller weight 2.0kg

 

Execution Unit
ZX01 ZX02 ZX03
Syringe 5μL~1000μL 10μL~60mL 0.5μL~100μL
Plage de débit 0.058nL/min ~ 1,587 ml/min 0.184nL/min~83.318mL/min 0.005nL/min~152.456μL/min
Canal 1~8 1~4 1~8
Max stroke of pump 70mm 110mm 70mm
Max rated thrust >2kgf,Adjustable thrust >10kgf,Adjustable thrust >1kgf,Adjustable thrust
Advance per microstep 0.198μm/μstep 0.156μm/μstep 0.1905μm/μstep
Linear speed 0.635μm/min~95.25mm/min 1μm/min~150mm/min 0.6096μm/min~91.44mm/min
Execution unit dimension 195×43.5×70mm 266×86×76mm
Execution unit weight 0.85 KG 1.8 KG 0.25 KG

 

Reference Table of Syringe Specifications and Flow Rate

ZX02

Syringe size Syringe ID (mm) Min flow rate (nl/min) Max flow rate (ml/min)
10μl 0.485 0.184 0.027
25μl 0.729 0.417 0.0626
50μl 1.03 0.833 0.125
100μl 1.457 1.667 0.250
250µL 2.304 4.169 0.625
500µL 3.256 8.326 1.248
1ml 4.699 17.342 2.601
5ml 11.989 112.89 16.933
10ml 14.427 163.469 24.520
20ml 19.05 285.027 42.754
30ml 21.59 366.090 54.913
60ml 26.594 555.459 83.318

ZX03

Syringe size Syringe ID (mm) Min flow rate (nl/min) Max flow rate (μl/min)
0.5µL 0.103 0.005 0.761
1µL 0.146 0.01 1.531
2µL 0.206 0.020 3.047
5µL 0.3257 0.051 7.618
10μl 0.485 0.112 16.893
25μl 0.729 0.254 38.166
50μl 1.03 0.507 76.190
100μl 1.457 1.016 152.456

Les paramètres de débit ci-dessus sont obtenus en utilisant un tube en silicone pour transférer de l'eau pure à température et pression normales., son utilisation réelle est affectée par des facteurs spécifiques tels que la pression, moyen, etc.. Ci-dessus pour référence seulement.

Dimension

Execution Unit Dimension (UN:mm)

Controller Dimension (UN:mm)

Déclaration:Le droit d’explication finale des informations ci-dessus appartient à Lead Fluid.

Leave Your Message