TFD02-01 Split Style Syringe Pump

Breve descripción:

Rango de flujo:0.184nL/min-83.318mL/min

Número máximo de canales: 1

 

Envíenos un correo electrónico

TFD02-01 Laboratory Syringe Pump

● Lead Fluid TFD02-01 laboratory syringe pump consists of two parts, controller and execution unit.exquisite structure, alta precisión, operación estable.

● El controlador y la unidad de ejecución están separados, a través de cables que conectan, y equipado con conector rápido, fácil de desmontar y montar. It can be used in
conjunction with other instruments and devices,como la estereotaxis cerebral.

● Execution unit can cooperate with 10μl~60ml syringe, which is suitable for various micro-injection

Característica

• Sistema de software LF-Touch-OS, funcionamiento estable y confiable, función poderosa.
• 4.3 inch color LCD touch screen operation, Se muestra la imagen de la animación y el icono., y colorida interfaz ,las teclas de función más utilizadas, hacer que el dispositivo sea más fácil de usar.
• A variety of working mode choice.
• Controller and execution unit split design, fácil de instalar y fijar.
• Support screen lock, operación de silencio de tecla.
• Button with indicator light, estado claramente de trabajo.
• Support a variety of standard syringe, personalizar la jeringa.
• High precision control.
• Comunicación RS485, Soporta el protocolo MODBUS..
• External signal control start-stop and direction.
• A wide range of power input.

Parámetro técnico

Work mode infuse, withdraw, infuse/withdraw, withdraw/infuse, continuous mode
Canal 1
Stroke of pump 110milímetros
Advance per microstep 0.156μm/μstep
Linear speed 1μm/min~150mm/min
Linear resolution 1μm/min
Linear travel accuracy error<±0.35%(>30% of pump stroke)
Max rated linear force >10kgf
Thrust regulation 1~100% arbitrary adjustable
Syringe size 10μL~60mL
Syringe selection built-in main manufacturers, the main model syringe to choose, can use custom syringe, direct input syringe size and diameter
Rango de flujo 0.184nL/min~83.318mL/min
Flow calibration through the calibration procedure to obtain more accurate fluid volume
Modo de visualización 4.3-inch color LCD, with screen display and transfusion volume, residual fluid volume, flow, operation display operation status,direction, syringe specification, Animation display operation status
Modo de operación touch screen + common function keys
Power-off memory storing the running parameters automatically
Function pause and stop,the sound tip,the lock parameters,fast forward and retreat quickly,display brightness adjustment
State signal output 1 road start-stop status ,1 road direction status
Control signal input 1 road the falling edge triggers start and 1 road the falling edge triggers stop
Communication RS485,protocolo modbus
Controller dimension 245×205×95mm
Execution unit dimension 266×86×76mm
Peso 2+1.8kg (controller + execution uni)
Fuente de alimentación CA100~240V,50/60Hz
Ambiente temperature 0~40℃,humedad relativa<80%

Flow Stability Curve

Syringe size Syringe ID (milímetros) Min flow rate (nl/min) Max flow rate (ml/min)
10μl 0.485 0.184 0.027
25μl 0.729 0.417 0.0626
50μl 1.03 0.833 0.125
100μl 1.457 1.667 0.25
250μl 2.304 4.169 0.625
500μl 3.256 8.326 1.248
1ml 4.699 17.342 2.601
5ml 11.989 112.89 16.933
10ml 14.427 163.469 24.52
20ml 19.05 285.027 42.754
30ml 21.59 366.09 54.913
60ml 26.594 555.459 83.318

Los parámetros de flujo anteriores se obtienen utilizando un tubo de silicona para transferir agua pura a temperatura y presión normales., su uso se ve afectado por factores específicos como la presión, medio, etc.. Arriba solo como referencia.

Dimensión (milímetros)