TYD01-02 Laboratory Syringe Pump

Kurzbeschreibung:

Durchflussbereich:0.184nL/min~83.318mL/min

Maximale Kanalanzahl: 2

Senden Sie uns eine E-Mail

Einführung

♦ Lead Fluid TYD01-02 adopts integrated table model design, 4.3 inch HD LCD touch screen display and operation, can assembly 2 plastic syringes or gas tight syringes, size range of syringe 10μL~60mL, flow range 0.184nL/min~83.318mL/min.
♦ Use excellent control system and precision mechanical structure, linear travel accuracy<±0.35%multiple working modes are available, powerful function,very suitable for high precision micro-transmission of various fluids in scientific research and experiment.
♦ Drive screw fixation using high strength aviation bearing, compared with traditional copper sleeve bearing, the service life of the screw has been significantly improved, and the accuracy is higher.
♦ Strong Anti-EMI performance, continuous and stable operation under high voltage electrostatic field.
♦ RS485 communication, compatibility MODBUS protocol, automation control system can be composed of computer, Plc, single chip computer etc. multiple host computers.

Funktion und Merkmal

• Assembly 2 plastic syringes or gas tight syringes.
• Multiple working mode options.
• Color LCD touch screen, convenient operation.
• Support screen lock, key mute operation.
• Button with indicator light,clearly working state.
• Support for a variety of syringes, can be customized syringes.
• High accuracy control.
• Syringe protection and traffic jam alarm.
• RS485-Kommunikation,support the MODBUS protocol.
• External signal contol start-stop and direction.
• Wide range power input.
• All-metal anti-static, anti-corrosion spraying process housing

 

Technischer Parameter

Work mode infuse, withdraw, infuse/withdraw, withdraw/infuse, continuous mode
Number of channe 2
Stroke of pump 110mm
Advance per microstep 0.156μm/μstep
Lineare Geschwindigkeit 1μm/min~150mm/min
Linear resolution 1μm/min
Linear travel accuracy error<±0.35%(>30% of pump stroke)
Max rated linear force >16kgf
Thrust regulation 1~100% arbitrary adjustable
Syringe size 10μL~60mL
Syringe selection built-in main manufacturers, the main model syringe to choose, Can use custom syringe, direct input syringe size and diameter
Durchflussbereich 0.184nL/min~83.318mL/min
Flow calibration through the calibration procedure to obtain more accurate fluid volume
Display mode 4.3-inch color LCD, with screen display and transfusion volume, residual fluid volume, flow, operation display operation status, direction,syringe specification, Animation display operation status
Operation mode touch screen + common function keys
Power-off memory storing the running parameters automatically
Function pause and stop,the sound tip,the lock parameters,fast forward and retreat quickly,display brightness adjustment
State signal output 1 road start-stop status ,1 road direction status
Control signal input 1 road start-stop status ,1 road direction status
Kommunikationsschnittstelle RS485,MODBUS-Protokoll
Dimension (L×B×H) 245×195×140mm
Gewicht 3.2kg
Stromversorgung AC100~240V,50/60Hz
Environment temperature 0~40℃relative humidity<80%

Reference Table of Syringe Specifications and Flow Rate

Syringe size Syringe ID (mm) Min flow rate (nl/min) Max flow rate (ml/min)
10μl 0.485 0.184 0.027
25μl 0.729 0.417 0.0626
50μl 1.03 0.8333 0.125
100μl 1.457 1.667 0.250
250μL 2.304 4.169 0.625
500μL 3.256 8.326 1.248
1ml 4.699 17.342 2.601
5ml 11.989 112.890 16.933
10ml 14.427 163.469 24.520
20ml 19.05 285.027 42.754
30ml 21.59 366.090 54.913
60ml 26.594 555.459 83.318

Die oben genannten Durchflussparameter werden durch die Verwendung eines Silikonschlauchs zur Übertragung von reinem Wasser bei normaler Temperatur und normalem Druck erreicht, Bei der tatsächlichen Verwendung wird es durch bestimmte Faktoren wie Druck beeinflusst, mittel usw. Oben nur als Referenz.

Dimension(UNmm)

Stellungnahme:Die endgültige Erklärung der oben genannten Informationen gehört zur Bleifluid.